The test is designed to diagnose menopause by detecting elevated levels of FSH in urine.

Accuracy > 99,9 %
Sensitivity > 99,9 %
Specificity > 99,9%

Results in 3 minutes

Sample - urine

The problem

Follicle-stimulating hormone (FSH) is important during the menopause because it reflects important changes in a woman's body: the ovaries become less sensitive to hormonal signals due to the increase in FSH. This change causes the typical symptoms of menopause and marks the end of the reproductive phase of a woman's life.

  • Current tests
  • Package leaflet

Wait until the urine sample has cooled to room temperature (15-30 °C) before testing.

  1. Set a day to take the test. (See section WHEN TO START TESTING).
  2. Keep the bag at room temperature before opening it. Remove the test from the sealed bag and use it within one hour.
  3. Remove the cap from the dipstick test and hold the test so that the urine stream flows over the absorbent tip, or dip the absorbent tip in
    (about 2/3) into a urine sample taken in a clean container and hold for at least 10-15 seconds.
  4. Put the cover on the current test and place the test on a flat and stable surface (the test and control box must face upwards).
    Switch on the timer immediately.
  5. Once the test has started, you may notice a light-coloured flow moving in the test and control window. Check the result after 3 minutes. Do not check the results after 10 min.
Positive Two lines are visible and the line in the test line area (T) is the same as or darker than the line in the control line area (C). A positive result means that the FSH level is higher than the normal level. Write down the result and refer to the table below to interpret the results.
Negative Two lines are visible, but the line in the test line area (T) is lighter than the line in the control line area (C) or there is no line at all in the test line area (T). A negative result means that the FSH level has not risen. Write down the result and refer to the table below to interpret the results.
Not valid The control line does not appear. The most likely reasons for the non-appearance of the reference line are insufficient sample quantity or an incorrect procedure. Revise the procedure and repeat the test using a new test. If the problem persists, stop using the test kit and contact your local distributor.

Ingredients

  • Current tests
  • Package leaflet

Instructions

Wait until the urine sample has cooled to room temperature (15-30 °C) before testing.

  1. Set a day to take the test. (See section WHEN TO START TESTING).
  2. Keep the bag at room temperature before opening it. Remove the test from the sealed bag and use it within one hour.
  3. Remove the cap from the dipstick test and hold the test so that the urine stream flows over the absorbent tip, or dip the absorbent tip in
    (about 2/3) into a urine sample taken in a clean container and hold for at least 10-15 seconds.
  4. Put the cover on the current test and place the test on a flat and stable surface (the test and control box must face upwards).
    Switch on the timer immediately.
  5. Once the test has started, you may notice a light-coloured flow moving in the test and control window. Check the result after 3 minutes. Do not check the results after 10 min.

Results

Positive Two lines are visible and the line in the test line area (T) is the same as or darker than the line in the control line area (C). A positive result means that the FSH level is higher than the normal level. Write down the result and refer to the table below to interpret the results.
Negative Two lines are visible, but the line in the test line area (T) is lighter than the line in the control line area (C) or there is no line at all in the test line area (T). A negative result means that the FSH level has not risen. Write down the result and refer to the table below to interpret the results.
Not valid The control line does not appear. The most likely reasons for the non-appearance of the reference line are insufficient sample quantity or an incorrect procedure. Revise the procedure and repeat the test using a new test. If the problem persists, stop using the test kit and contact your local distributor.

Instruction leaflet

You can buy this test:

Back to top Drag